首先祝你们圣诞节快乐。
其次,就泵的这个问题我觉得双方还是有必要再协商。迄今为止已经发生了不止5次吸样泵的损坏。造成了漏水,导致了电路板被腐蚀。不得不更换新的泵。(此类情况一般发生仪器使用一年左右)没有任何一台仪器厂家的泵是属于耗材的,而且又会经常坏,这个太不可思议的。
我们认为这是泵的本身的质量问题,和我们的安装使用,并无多大的关系。由于我们安装了大量的仪器在中国,而且按照这个进度,我估计大概在全中国会有近30%的仪器会发生类似的问题,这给我们仪器造成很不好的影响。所以我再次建议可以免费给我们更换所有的泵。
First of all I wish you merry Christmas。
Second, the problem of the pump I think both sides still necessary to negotiate。 So far has happened more than five times the damage of the pump suction samples。 Caused the leak, lead to a circuit board corrosion。 Have had to change the new pump。 (such ...全部
First of all I wish you merry Christmas。
Second, the problem of the pump I think both sides still necessary to negotiate。
So far has happened more than five times the damage of the pump suction samples。 Caused the leak, lead to a circuit board corrosion。
Have had to change the new pump。 (such generally occur instrument used in a year or so) no one instrument of the pump manufacturer is to belong to the consumables, and will often bad, this is incredible。
We think that this is the pump itself the quality question, and we use the installation, and does not have how old relationship。
Because we installed a large number of instruments in China, and according to the schedule, I estimate that in all the Chinese will have nearly 30% of the instrument can produce similar problems, this gives us cause is not the shadow of the instrument
。